10°C To 25°C 66L/Min 60L Capacity Recirculating Chiller

Across International

10°C To 25°C 66L/Min 60L Capacity Recirculating Chiller

The WAC-6 Digital Recirculating Chiller offers precise cooling from 10°C to 25°C with a 66L/min flow rate and 60L capacity. Featuring a durable stainless steel tank, water pump, and included water manifold, it’s built for reliable, rust-free performance. Perfect for IH series induction heaters and lab setups!

$13,886.55
$9,590.00USD

10°C To 25°C 66L/Min 60L Capacity Recirculating Chiller

Stay cool under pressure with the WAC-6 Digital Recirculating Chiller, engineered for precision and durability. Designed to maintain optimal temperatures between 10°C and 25°C with an impressive 66L/min flow rate and 60L capacity, this chiller is the ideal companion for IH series induction heaters and demanding lab applications.

Built for High Performance

  • Durable Design: Stainless steel tank, water pump, and tubing ensure rust-free, long-lasting operations.
  • Precision Cooling: Digital temperature control delivers consistent performance for sensitive applications.
  • Plug-and-Play: Comes with a water manifold for immediate use.
  • Convenient Mobility: Built-in swivel casters make it easy to transport and position.

Whether you’re working in a lab or industrial setting, the WAC-6 Recirculating Chiller is the ultimate cooling solution for high-performance, hassle-free operations. Upgrade your system today!

Reliable. Efficient. Built to Perform.

Author

Product Specifications

Specification Details
Model WAC-6
Controller Digital micro-chip controlled with LCD screen and fluid level display
Input Power 480V 60Hz three-phase (3PH) 5.5kW
Nominal Cooling Capacity 14028 kcal per hour
Nominal cooling capacity is according to:
– Inlet chilled fluid temperature: 12°C
– Outlet chilled fluid temperature: 7°C
– Inlet cooling air temperature: 30°C
– Outlet cooling air temperature: 35°C
Water Pump 480VAC 50/60Hz three-phase (3PH) 750W
3450 rpm. Max pressure 10 bar. Max liquid temperature: 70°C
Pumping speed: 66L per minute
Reservoir Capacity 60L
Recommended Coolant Distilled water
Compressor 460VAC 50/60Hz three-phase (3PH) 10A Max
Thermally protected. PS: 22.6 bar
TS max: 50°C, TS min: -35°C
Lubricant: P.O.E. 160 PZ
Refrigerant: R407C-R134a-R404A-R507
Temperature Accuracy +/- 2 °C
Condenser Cooling Air Flow 212,000 cu ft per hour
Chilled Fluid Flow 60L/min
Inlet & Outlet Diameter 25mm/1 inch (four inlets and four outlets)
Unit/Shipping Weight 200 kg / 260 kg
Unit/Shipping Dimensions 1040 x 610 x 1370 mm / 1220 x 660 x 1520 mm (DxWxH)
Safety Over current, compressor delay start, water pump overload, built-in fuse, over-voltage and under-voltage, abnormal operations, phase protection
Working Environment Temperature range of chilled fluid: 5°C – 35°C
Temperature difference of chilled fluid between inlet and outlet: 3°C – 8°C
It is better to use the chiller while the ambient temperature is at or below 35°C
Recommended Induction Heaters High frequency: 20kW, 30kW, 40kW
Mid frequency: 35kW, 45kW
Low frequency: 35kW, 45kW
CE Compliant Yes